TECHNICAL FEATURES



FLOW RATE: 40 SLPM 40 SLPM 100 SLPM


PURITY: 98% to 99.9%


TECHNOLOGIES EMPLOYED: PSA


OUTPUT PRESSURE: 0 to 7 barg / 15 to 100 psig


OXYGEN AT MAX FLOW: < 2%


MOISTURE AT MAX FLOW: - 44°C dew-point STP, < 80 ppm


AIR INPUT CHARACTERISTICS: 

Particulate ≤ 0.01 microns


Oil vapors ≤ 0.01 mg / m3


Moisture ≤ 3°C dew point in pressure


Min pressure 8.5 barg / 120 psig


Max pressure 10 barg / 145 psig


AIR COMPRESSOR: No for NiGen LCMS 40.0 and  NiGen LCMS 100.0 - Yes for NiGen LCMS 40.1


ELECTRICAL SUPPLY: 

115 VAC ± 10%, 60 Hz


230 VAC ± 10%, 50 Hz


INDEX OF PROTECTIONIP20


OPERATING TEMPERATURE: 40°F to 100°F / 5°C to 40°C


ENVIRONMENTAL MOISTURE: ≤ 80%


MAX ALTITUDE: 2000 m or 6560 ft above sea level


DIMENSIONS: 

51.1”(H) x 17.3” (W) x 43.3” (D)


130 cm (H) x 44 cm (W) x 110 cm (D)


WEIGHT: 198 lbs / 90 kg 220 lbs / 100 kg 330 lbs / 150 kg


APPLICATIONS: LC-MS E, sample preparation



ADDITIONAL FEATURES


• Air + N2 configuration for AB LC-MS


• High efficiency hydrocarbon removal columns included


• Air + N2 configuration for Agilent Jet Stream LC-MS


• CPU for control of up to 32 units


• CAN-BUS communication


• Maintenance diary


• Report alarms


• Equipped with wheels



ACCESSORIES


• AB pressure regulator kit


• UHP available nitrogen purifier for Agilent collision cell on Jet Stream LC-MS


• Inlet air coalescing filter


• 3 point pressure regulator kit


• Installation kit


• Automatic 2 sources switch kit